SPECIFICATIONS |
|
INSULATION RESISTANCE |
|
Range |
Resolution |
Accuracy |
Output Voltage, DCV |
30 ?C 1000 |
1 |
?? (1% of setting + 1 V) |
Output Regulation DCV |
?? (1% of output + 1 V) from no load to full load |
Output Ripple |
< 2 % |
Discharge |
Automatic Discharge of Device Under Test |
Indicator : Green < 30 V, Red > 30 V |
SETTINGS |
Low-Limit Resistance, M?? |
0.1 ?C 999.9 |
0.1 |
Same as Resistance Measurement |
1000 ?C 9999 |
1 |
10.0 G???C 50.0 G?? for MODEL 8205 |
0.1 G?? |
Delay Time, second |
1 - 999.9 |
0.1 |
??(0.1% + 0.05 sec) |
Dwell Time, second |
0, 0.1 ?C 999.9 (0=continuous) |
MEASUREMENT |
DC Voltage, V |
0.0?C100.0 |
0.1 |
?? (1% of reading + 1 V) |
101?C1000 |
1 |
Resistance |
0.001M?? - 50.0G?? for MODEL 8205 0.001M?? - 200.0G?? for MODEL 8210, Auto Range |
0.001M?? 0.01M?? 0.1M?? 1M?? 0.01G?? 0.1G?? |
500?C1000V |
0.1M?? - 1G??,?? (2% of reading + 2 counts) |
1G?? - 20G?? , ?? (3% of reading + 2 counts) |
20G?? ?C50G??,??(10% of reading + 2 counts) for MODEL 8205 |
20G???C200G??,??(10% of reading + 2counts) for MODEL 8210 |
30?C499V |
0.1M??-1G??,?? (3% of reading + 2 counts) |
1G?? - 20G?? , ?? (5% of reading + 2 counts) |
GENERAL |
Input Voltage AC |
115/230Vac??15%, 50/60Hz ?? 5%, Fuse 1A/ Fast acting 250VAC |
PLC Remote Control |
Inputs : Test , Reset, Interlock |
|
Outputs : Pass, Fail , Progressing |
Display |
4 Digits LED Display |
Terminations |
High Voltage Output ?C Alden Socket |
Shielded Return ?C BNC Connector |
Safety |
Built-in Smart GFI circuit,GFI trip current 450?A max, HV shut down speed: <1mS |
Calibration |
Built-in software and external calibrated meters |
Environment |
0-40??, 20-80%RH |
Dimension/Net Wight |
280mm(W)??89mm(H)??300mm(D)/ 4.5Kg |
STANDARD ACCESSORIES |
Power Cord(10A) |
??1 |
Fuses |
??2 (Including a spare contained in the fuse holder) |
Interlock Disable Key(1505) |
??1 |
Hipot Test Lead ,1.5m(1101) |
??1 |
BNC Connector Return Test Lead ,1.15m(1140) |
??1 |
*product specifications are subject to change without notice. |